![background image](http://html.mh-extra.com/html/somat/edaqlite/edaqlite_user-manual_1317281139.webp)
I2773-4.
4
en
139
Somat eDAQ
lite
ASCII File
Specify the full path name of the ASCII file that defines the mapping conditions. Use
the Browse button to select the desired file. Use the Check File option to parse the
ASCII file and verify that the format is valid.
NOTE
Because TCE parses the ASCII file at test initialization, any changes to the file prior to
initialization are effective for subsequent test runs.
Use Default Output
Select the default output option to output a default value when the input channel does
not meet any of the mapping conditions. If not selected, the output remains in its
existing state.
Default Out
Specify the default value for the output. The channel outputs the value if the first input
sample does not meet any mapping conditions. Also, when using default out, the
channel outputs the default value throughout the test run when the input does not meet
any mapping conditions.
Latch Period
Specify the time in seconds that the input channel must consistently map to the same
output state before the output state switches. The latch period is similar to a duty cycle
on the output state preventing the output state from switching for at least this period of
time. This feature can eliminate state switching transients in the output channel data
stream. If the latch period is 0.0, then the output state switches on each sample.
7.2.8
Statistical Analysis
The Statistical Analysis channel generates statistical output data from source
transducer or computed channel input data.
Input Channel
The input channel data type must be 32-bit float.
Output Data Type
The output data type is 32-bit float.
Statistic
Select one of six available statistical modes. The following table defines the algorithms
used to compute the statistical values where
N
is the number of data samples in the
analysis window and
X
mean
is the mean of the data in the analysis window.
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...