Leadwire Resistance Correction
Select the leadwire resistance correction option to compensate for leadwire resistance
effects when using the defined span or external span calibration modes.
NOTE
The eDAQ
lite
always performs leadwire resistance correction for shunt calibrations
and transducers calibrated using the defined span or external span calibration modes.
Leadwire Resistance
Specify the value of the leadwire resistance when using the leadwire resistance
correction option. The resistance input is the resistance of one lead, ideally measured
from the SMART module connector pin to the connection at the active bridge leg. It is
presumed that all lead wires are approximately the same length. Quarter-bridge
applications require the use of all three wires.
To accurately measure the leadwire resistance, use the four-wire resistance
measurement method. Nominal resistance values for transducer cable wires can help
to estimate the leadwire resistance (see
“Cable Resistances” on page 223
). However,
since contact resistances at mechanical connections are difficult to estimate, measure
the leadwire resistance for optimum accuracy.
Alternatively, TCE can back calculate the leadwire resistance if the bridge resistances
are equal and known and if the resistances of the three relevant leadwires are equal.
To use this method, the bridge type and bridge resistance fields must be correct.
Ensure the transducer is stable and then click the back calculate resistance option.
Back Calculate Resistance
Select the back calculate resistance option to experimentally measure the leadwire
resistance. Note that the measurements are only valid when the bridge resistances
are equal and all leadwire resistances are equal. If the resulting values show
significant difference between the upscale and downscale resistance values, the three
relevant leadwires are not equal and the back calculate method is invalid.
Resistor to Shunt Across
Select either the upscale (-Sig to -Ex) shunt option or the downscale (-Sig to +Ex)
shunt option. Note that the downscale shunt option results in a negative span signal
value.
NOTE
For older SMART bridge modules, the resistor to shunt across is fixed at production.
Hardware Configuration
Click the hardware button to view the SMART module user data parameters as they
are defined in the hardware setup configuration. Note that TCE does not update
reprogrammed SMART modules until a hardware query is performed.
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...