3. Click on the hardware tab and then Manage Network Cameras beneath the
hardware table to navigate to the camera setup page.
4. Click the Find Network Cameras button under Available Networked Cameras to
add the networked camera. The new entry is named “Misconfigured Camera.”
Figure 5-7:
Example entry for an unconfigured camera labeled as “Misconfigured Camera.”
5. Verify that the serial number for the new entry matches the serial number on the
shell of the camera. If it does not match, follow the troubleshooting steps in the next
section.
6. Click on the correct camera configuration link under the camera entry to open the
camera configuration page. Follow the on screen instructions to assign the camera
an IP address.
7. Wait during the configuring camera screen for the camera to be properly configured
and the web interface to return to the manage network cameras page. If the
configuration fails, try power cycling any switches or routers on the network. If
problems persist, refer to the troubleshooting steps in the next section.
Figure 5-8:
Example entry for a properly configured camera.
8. Click the select this camera link under the camera entry to add the camera to the
Selected Network Cameras list. (If there is no camera under Available Networked
Cameras click the Find Cameras on Network button.)
Figure 5-9:
Selected Networked Cameras table showing the cameras to be added to the
eDAQ
lite
hardware list.
9. Click the update eDAQ
lite
hardware list to reset the eDAQ
lite
and add the camera
to the eDAQ
lite
hardware table on the main hardware page.
NOTE
Once configured, a camera my be used by any networked eDAQ or eDAQ
lite
simply
by following steps 8 and 9 for each system. However, the camera cannot provide the
maximum frame rate to more than one eDAQ system at a time. For example, two
eDAQs using the same camera and configured to request 30 frames per second only
receive a maximum of 15 frames per second each.
HBM: public
Содержание eDAQlite
Страница 1: ...User Manual English Deutsch Espanol I2773 4 4 en HBM public Somat eDAQlite with TCE Software ...
Страница 2: ...Somat eDAQlite 2 I2773 4 4 en HBM public ...
Страница 16: ...Somat eDAQlite 16 I2773 4 4 en HBM public ...
Страница 22: ...Somat eDAQlite 22 I2773 4 4 en HBM public ...
Страница 88: ...Somat eDAQlite 88 I2773 4 4 en HBM public ...
Страница 106: ...Somat eDAQlite 106 I2773 4 4 en HBM public ...
Страница 128: ...Somat eDAQlite 128 I2773 4 4 en HBM public ...
Страница 190: ...Somat eDAQlite 190 I2773 4 4 en HBM public ...
Страница 202: ...Somat eDAQlite 202 I2773 4 4 en HBM public ...
Страница 222: ...Somat eDAQlite 222 I2773 4 4 en HBM public ...
Страница 224: ...Somat eDAQlite 224 I2773 4 4 en HBM public ...
Страница 227: ...I2773 4 4 en 227 Somat eDAQlite HBM public ...