7 Appendices
Sirona Dental Systems GmbH
Error processing
Operator’s Manual SIDEXIS XG
Identification
These messages are usually identified by an error number enclosed in
brackets in the title line of the message and the following symbol:
Below you will find a list of some of the error messages generated by SIDEXIS
XG with information for the user how to eliminate the problem.
See Section "List of error messages [
7.4.2
List of error messages
Explanation
The list describes some of the error messages generated by SIDEXIS XG.
Information how the user may eliminate the problem is also provided.
Error list
NOTICE
Depending on the frequency and effects of this message, it may be
necessary to call a service engineer.
NOTICE
If this information does not lead to the desired result or the error is not
included in the list, call the hotline for assistance.
1500
SIDEXIS can only be started once at each station. Limit
yourself to one instance of the software during your work.
1501
Possibly insufficient working memory (RAM) when starting
SIDEXIS. Exit other programs, if possible.
1502
Possibly insufficient working memory (RAM) when starting
SIDEXIS. Exit other programs, if possible. It should be
possible to continue working without the hotkeys, though.
1530
Check the connections, cables and connectors between
the PC, SIROCAM and the foot switch. It should be
possible to continue working without the camera and foot
switch, though.
1599
Reboot the PC. If the error occurs repeatedly, contact the
hotline.
1600
Insufficient working memory (RAM). Close any images no
longer needed or exit other programs, if possible.
1601
see 1600