9
Checking the system adjustment
Sirona Dental Systems GmbH
9.1
Test exposures
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
84
D 3352.031.03.18.02
9.1.1 Test exposure menu
The
panoramic X-ray unit
is already completely adjusted
on delivery. Check the adjustment using the
TEST
EXPOSURE
menu to make sure that the unit was not inad
-
vertently deadjusted during transport.
The
TEST
EXPOSURE
menu is started from
SIDEXIS XG
:
UTILITIES
CONSTANCY
TEST
XCXP
SELECT
X
-
RAY
DEVICE
SERVICE
EXPOSURE
SELECT
X
-
RAY
COMPONENT
DIAPHRAGM
TEST
EXPOSURE
IMPORTANT
The
SELECT
X
-
RAY
DEVICE
and
SELECT
X
-
RAY
COMPO
-
NENT
dialogs appear only if several x-ray devices or x-ray
components are available for selection.
IMPORTANT
When you open the
TEST
EXPOSURE
menu, the unit
switches from the user mode to the PC service mode
logged by the PC. The service mode is indicated by the
SERVICE
display on the Multipad (
2.
).
In the
PC service mode
, the control options
that are available on the Multipad are determined by SI
-
DEXIS and the currently selected service routine. General
control of the unit by means of the Multipad (as in the user
mode) is not possible in this mode.
The test exposures are used to check the adjustment. They
can be started without a password.
You can switch between the
PAN
DIAPHRAGM
and
PAN
SYMMETRY
menus by clicking the corresponding
tab
with
the mouse. To quit the
TEST
EXPOSURE
menu, click
CANCEL
.
Displays on the Multipad
During the adjustment procedure, different service routines
are started from SIDEXIS; they are displayed one after the
other on the Multipad.
If there are any error messages displayed during adjust
-
ment, please follow the instructions provided in the Service
Manual.
9.1 Test exposures
2.
S E R V I C E
1.
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......