9
Checking the system adjustment
Sirona Dental Systems GmbH
9.1
Test exposures
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
88
D 3352.031.03.18.02
9.1.4 Checking the PAN symmetry
WARNING
When performing the following tests, be sure to ob
-
serve the radiation protection regulations applicable in
your country (see Operating Instructions).
•
Insert the needle phantom in the bite block holder of the
panoramic X-ray unit (see page 85).
IMPORTANT
The sensor must be plugged into the sensor slot on the pan
-
oramic X-ray unit.
1.
Go to the
PAN
-
SYMMETRY
submenu.
2.
To make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
The exposure dialog box showing the exposure status
appears in Sidexis.
The initialization status is visualized by a progress indi
-
cator on the Multipad.
The initialization procedure is completed when the ex
-
posure parameters of service routine
S010.2
(60 kV / 3 mA; 14.1 s) are displayed and the progress
indicator disappears.
3.
Take an exposure (60 kV/3 mA):
– Press the
R key
on the Multipad to move the unit
back to the starting position.
– Press the release button. Hold down the release
button until image acquisition is completed and the
preview image appears in the exposure dialog box.
1.
2.
in SIDEXIS
3.
T
R
Prog.
S
kV
mA
G
on the Multipad
6 0 / 3
1 4 . 1
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......