Installation Instructions ORTHOPHOS XG 3
11.2
Adjusting the panoramic X-ray unit
60 51 416 D 3352
D 3352.031.03.18.02
111
11.2.2Important information concerning adjustment
WARNING
When performing the following tests, be sure to ob
-
serve the radiation protection regulations applicable in
your country (see Operating Instructions).
WARNING
“Radiation” is signaled by an X-ray LED and a beep.
IMPORTANT
Be sure to take screenshots of the
PAN
-
RESET
ADJUST
-
MENT
menu before and after the adjustment (see section
11.2.8) and save them to the C:\SIDEXIS\XGRAW directory
along with the time and date!
IMPORTANT
Before starting the service routines for system adjustment,
make sure that no unit movements are active (especially di
-
aphragm travels)! Otherwise the system may become inop
-
erable in rare cases.
Coarse and precision adjustment
using the Diaphragm/system
adjustment menu in SIDEXIS XG
IMPORTANT
The Pan - Sensor adjustment submenu contains a coarse
adjustment and a precision adjustment (precision adjust
-
ment is preset). Always try to use precision adjustment first
when adjusting the sensor. In most cases, previous coarse
adjustment is not necessary.
With SIDEXIS Version V02.20 and higher, a message win
-
dow indicates when a coarse adjustment is necessary.
Only if you cannot achieve your goal with precision adjust
-
ment, e.g. if the exposed area is completely outside the im
-
age field, should you perform a coarse and then a precision
adjustment.
If a coarse adjustment proves necessary, deactivate the
PRECISION
ADJUSTMENT
check box and follow the adjust
-
ment steps described in the present chapter to perform a
coarse adjustment.
The steps and correction procedure required for coarse ad
-
justment are identical to those for precision adjustment. The
only difference between the two modes is the size of the im
-
age section considered. Furthermore, there are fewer auxil
-
iary lines in the coarse adjustment mode.
Default values in the Diaphragm/system adjustment
menu in SIDEXIS XG
During the adjustment, the default adjustment values are
displayed in the text boxes of the Diaphragm/system adjust
-
ment menu.
First perform the adjustment with these default values. If
you do not attain the desired result via this automatic
adjustment, you should determine the adjustment values
manually by measuring the exposure with the SIDEXIS
measuring ruler and then overwrite the default values in the
menu.
This procedure is described in the following sections.
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......