11
Appendix
Sirona Dental Systems GmbH
11.2
Adjusting the panoramic X-ray unit
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
116
D 3352.031.03.18.02
Image:X-ray image with unadjusted pan sensor:
Alternatively with precision adjustment setting B and
coarse adjustment setting A
Coarse or precision adjustment
?
Sensor adjustment can usually be performed directly via
precision adjustment. Only in exceptional cases, e.g. if one
or more needles are completely outside of the image field
(A)
in an image acquired with the
PRECISION
ADJUSTMENT
presetting, is it necessary to perform a coarse adjustment
prior to precision adjustment
(B)
. To do this, deactivate the
PRECISION
ADJUSTMENT
check box (see page 111) and
then perform a coarse adjustment proceeding in the same
way as for precision adjustment. The only difference
between coarse and precision adjustment is the size of the
image area considered. Furthermore, there are fewer aux
-
iliary lines in the coarse adjustment mode.
IMPORTANT
With SIDEXIS SW V01.45 and higher, a message window
indicates whether a coarse adjustment is required on com
-
pletion of the precision exposure.
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......