Manual adjustment of the PAN sensor
The manual adjustment procedure is similar to the one for
automatic adjustment. The only difference is that the default
adjustment values automatically determined by SIDEXIS
are overwritten by manually determined adjustment values
in the
PAN
-
SENSOR
ADJUSTMENT
submenu.
1.
Measure distances
S1
,
S2
and
S3
with the SIDEXIS
measuring ruler.
IMPORTANT
To measure S1, S2 and S3, estimate the horizontal center
position of the displayed needles. Measure in the lower area
of the needles if possible, since they may have been bent
slightly after repeated use.
Tip:
To facilitate the measuring procedure, you can color
the image in SIDEXIS (see also SIDEXIS Operator’s Manu
-
al).
2.
Overwrite the default values for
S1, S2
and
S3
with the
measured values in the text boxes of the
PAN
-
SENSOR
ADJUSTMENT
submenu.
IMPORTANT
For information on the direction of displacement (input of +/
- sign in the menu) see page 110. Use points as decimal
separators!
•
Proceed with step 5 of the adjustment procedure on
page 117.
1.
S1
S2
S3
2.
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......