Sirona Dental Systems GmbH
11
Appendix
Installation Instructions ORTHOPHOS XG 3
11.2
Adjusting the panoramic X-ray unit
60 51 416 D 3352
D 3352.031.03.18.02
117
IMPORTANT
The default values for S1, S2 and S3 were automatically de
-
termined by SIDEXIS based on the exposure and entered in
the text boxes of the menu.
For manual adjustment, the values displayed at this position
in the text boxes of the menu can be overwritten.
First continue with the automatic adjustment. Manual deter
-
mination of the adjustment values is required only if you fail
to reach your goal via automatic adjustment (see page 120).
5.
Make SIDEXIS XG ready for exposure:
Click
IMAGE
ACQUISITION
The
PAN
-
SENSOR
ADJUSTMENT
dialog box appears in
Sidexis.
The dialog box suggests a value
Lx
for the mechanical
adjustment of the sensor.
If the suggested value is greater than ±0.5 mm, then
perform a mechanical adjustment of the sensor.
6.
Adjust the sensor:
IMPORTANT
Positive sign =
Moves the sensor to the right
Negative sign =
Moves the sensor to the left
To do this, proceed as follows:
– Mark the current position of the sensor with a per
-
manent marker.
– Loosen the three screws
A
(approx. 2 turns, do not unscrew fully)
– Move the sensor to the left or right by the displayed
value
Lx
.
Tighten screws
A
again.
5.
System version 1: Fixed sensor holder
6.
A
A
A
+
-
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......