7
Initial startup
Sirona Dental Systems GmbH
7.4
Checking the data paths
Installation Instructions ORTHOPHOS XG 3
60 51 416 D 3352
70
D 3352.031.03.18.02
3.
Select/confirm the X-ray device:
Select e.g.
ROOM
3
and click
OK
.
IMPORTANT
The
SELECT
X
-
RAY
DEVICE
dialog appears only if several
X-ray devices are installed.
The dialog box for selecting the test type appears on
the screen.
4.
Select/confirm the test type:
Click
SERVICE
EXPOSURE
.
The dialog box for selecting the service exposure ap
-
pears on the screen.
5.
Select/confirm the service exposure:
Click
DIGITAL
TEST
PATTERN
– If several different X-ray components are available,
a dialog box for selecting the X-ray component
appears on the screen. In this case, select/confirm
the required component.
– If only one X-ray component is available, the expo
-
sure readiness dialog box appears on the screen
and shows the status of the exposure.
IMPORTANT
During operation in the service mode, the unit switches from
the user mode to the PC service mode logged by the PC.
The service mode is indicated by the
SERVICE
display on
the Multipad (see page 84).
The unit switches back to the user mode as soon as the ex
-
posure is completed.
3.
5.
4.
Содержание orthophos xg 3
Страница 1: ...NMKOMNR båÖäáëÜ kÉï ë çÑW loqelmelp ud P fåëí ää íáçå fåëíêìÅíáçåë T R Prog S kV mA P1 ...
Страница 5: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 5 1 Before you begin ...
Страница 13: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 13 2 Delivery and transport ...
Страница 19: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 19 3 Installation Panoramic X ray unit ...
Страница 45: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 45 4 Electrical connection ...
Страница 49: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 49 5 Installation Remote control ...
Страница 59: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 59 6 Safety checks ...
Страница 65: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 65 7 Initial startup ...
Страница 73: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 73 8 Startup for USA Canada only ...
Страница 83: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 83 9 Checking the system adjustment ...
Страница 91: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 91 10 Final work ...
Страница 97: ...ORTHOPHOS XG 3 60 51 416 D 3352 D 3352 031 03 18 02 97 11 Appendix ...
Страница 141: ......