Sirona Dental Systems GmbH
Starting up
Installation Instructions HELIODENT
PLUS
Functional test
båÖäáëÜ
7
Starting up
7.1
Functional test
7.1.1
Overview
empty
Explanation
A functional test must be performed according to the following procedure.
empty
Step 1
1. Check the protective ground wires and the device leakage current
according to IEC 62353: 2007 (see section "Protective conductor test
[ ➙ 54]" and "Leakage current test [ ➙ 56]“).
Record the measured values in chapter 3 of the document
"Inspection, maintenance and safety-related check".
2. Connect the power cable now.
empty
Step 2
➢ Check the function of the X-ray tube assembly (see Section
"Checking the X-ray tube assembly [ ➙ 57]“).
empty
Step 3
At least one of the following tests must be performed, depending on which
measuring instruments are available:
➢ Check the tube current (see the section entitled "Checking the tube
or
➢ Review the exposure time and high voltage kV (see Section
"Checking exposure time and high voltage kV [ ➙ 60]").
empty
Step 4
➢ Check the deadman function (see section "Deadman function of the
Содержание Heliodent plus