Sirona Dental Systems GmbH
Electrical connection
Installation Instructions HELIODENT
PLUS
Connecting the control panel
båÖäáëÜ
6.3.2
Connection to the remote timer
empty
Scope
For installation options: 3, 5, 7 and 8
empty
Connection
1. Plug cable L6 into the connecting strip X400 of PC boardDX1.
2. Secure cable L6 with the shielding clamp (D).
3. Connect cable L6 to strips X105 and X106 of PC board DX4.
4. Secure cable L6 with the shielding clamp (V).
NOTICE
The shielding of cable L6 must be in contact with the shielding clamp
(D).
X106.1
X106.2
X106.3
X106.4
X105.1
X105.2
X105.3
X105.4
WH
GY
BN
GN
YE
PK
BU
RD
NOTICE
The shielding of cable L6 must be in flat contact with the shielding clamp
(V).