Starting up
Installation Instructions HELIODENT
PLUS
Attaching the housing covers
båÖäáëÜ
7.2.4
Remote timer
1. Only with manual release S3 (coiled cable): Place the ferrite core of
the coiled cable inside the housing centered on the opening in the
wall. To do this, fasten the coiled cable with the strain relief clamp (S).
2. Mount the cover of the Remote Timer on top of the chassis from
above.
3. Snap the cover into the chassis at the bottom.
4. Only with manual release S3 (coiled cable):Cover the release symbol
on the control membrane with tape.
5. Only with manual release S3 (coiled cable): Hang the manual release
into the holder next to the control membrane.
S
NOTICE
The blank panel must be firmly attached to the cover on the wall module.