62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
13
Sirona Dental Systems GmbH
Required materials
Installation Instructions HELIODENT
PLUS
Required installation material
båÖäáëÜ
4
Required materials
4.1
Required installation material
Wall properties
● The permissible tensile force of the selected attachment must at least
equal the tensile force listed above.
● Matching wood screws for wooden beams are included in delivery.
● For all other wall structures, special wall anchors must be purchased
from a selected dealer. The wall anchors and screws should be
identical for every attachment point.
● Alternatively, an anchor plate can be used as a counter bearing. In
this case, M8 threaded rods of the appropriate length for the wall
(thickness of the wall + 2 x mounting plate thi attachment
material) are required.
CAUTION
Observe wall properties
In installation situations, the technician is responsible for the
assessment of wall properties and selecting the method of attaching the
unit to the wall.
K
Tensile force per screw
3600 N (800 lbf) if L ≤ 700mm (27 1/2“)
4200N(950 lbf) if L ≤ 950mm (37 3/8“)
L
Length of support arm
M
Mounting plate (supplied)
N
Anchor plate
O
Threaded bolt M8
K
L
M
N
O