62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
17
Sirona Dental Systems GmbH
Installation
Installation Instructions HELIODENT
PLUS
Installing the wall adapter
båÖäáëÜ
5.1.2
Installation (concealed installation)
1. Remove wall adapter (P1) from the packaging.
2. For the power cable: Carefully punch out opening B in the wall
adapter with a punch.
3. Position the wall adapter on the prepared installation location.
4. Pull the power cable through the opening (B) in the wall adapter.
5. For remote control / Remote Timeronly (cable L6 or L9): Pull the
cable through the opening (F) in the wall adapter.
6. Screw the wall adapter onto the wall loosely to hold it in place.
7. Align the wall adapter with a spirit level and attach the wall adapter
securely with suitable screws and washers, depending on the type of
attachment.
P1
CAUTION
Risk of damage
Make sure that board DX1 is not damaged during installation!
Observe ESD protective measures!
L1
PE
N´
N
B
F
L9/L6
Содержание Heliodent plus
Страница 1: ...NNKOMNM ebiflabkqmirp k W b f f HELIODENT PLUS 7mA DC 7mA DC 70kV 60kV S 008...
Страница 2: ......