Electrical connection
Sirona Dental Systems GmbH
Connecting the release and the door contact
Installation Instructions HELIODENT
PLUS
6.4.9
Installation option 6.1
empty
Configuration
Installation option 6.1
Release outside of the X-ray room with remote control, door contact
safety circuit
● Door contact
– Door contact S7 wired to the remote control housing
● Release
– Manual release S3
or
– Remote control release key S9
As in installation version 6:
empty
Exception
Door contact wired via remote control.
empty
Prerequisites
● The remote control has been installed (see section "Installing the
remote control/remote timer [ ➙ 23]").
● Only with use of the manual release S3 (coiled cable):
– The holder for manual release S3 was attached to the cover of
the remote control. See section "Installation on the protective
cover of the remote control [ ➙ 29]".
6.1
L9
S7
S3
S9
NOTICE
Installation requirement
Use of the remote control is permissible only if the yellow X-Ray LED is
visible to the operating personnel during radiation release.