62 14 972 D 3507
D 3507
.
031.01.07
.
02 11.2010
9
Sirona Dental Systems GmbH
Installation options
Installation Instructions HELIODENT
PLUS
båÖäáëÜ
2
Installation options
empty
Designations for release buttons and door contact
Designations of release buttons and door contact
● Manual release S3
– Coiled cable
● Release key on the control membrane S4
– Directly connected to control board DX4
● Remote control release key S9
– Integrated in remote control housing
● Door contact (safety circuit) S7
empty
Installation option 1
Installation option 1
Release in the treatment room without remote control
● Release
– Manual release S3
empty
Installation option 2
Installation option 2
Release in the treatment room with remote control
● Release
– Manual release S3
or
– Remote control release key S9
empty
Installation option 3
Installation option 3
Release in the treatment room with Remote Timer
● Release
– Manual release S3
or
– Release key on the control membrane S4
1
S3
2
L9
S3
S9
3
L6
S3
S4
NOTICE
Length of cable supplied for Remote Timer approx. 10 meters (394")
(must not be extended).
Conduit int. dia. at least 12 mm (1/2").
Содержание Heliodent plus
Страница 1: ...NNKOMNM ebiflabkqmirp k W b f f HELIODENT PLUS 7mA DC 7mA DC 70kV 60kV S 008...
Страница 2: ......