62 14 972 D 3507
72
D 3507
.
031.01.07
.
02 11.2010
Appendix
Sirona Dental Systems GmbH
Replacing old units
Installation Instructions HELIODENT
PLUS
Preparation
1. Unscrew and remove the old unit.
2. Check the existing drill holes. Do they comply with the description in
the installation regulations?
3. Check the electrical connection area of the old unit. It may be
necessary to relocate the electrical connections (concealed
installation for example).
4. Only if the old power cable is used: Check whether the minimum
requirement for the power cable (3x1.5 mm
2
(AWG 16)) has been
fulfilled.
5. Only for concealed installation of a remote control or a remote timer:
Check whether a conduit with a ⌀ int. of at least 12mm (1/2“) is
available for cable L6 or L9.
empty
Concealed installation
1. Only if the old power cable is used: Extend the power cable
sufficiently if necessary.
2. Only if a remote timer is used: Pull cable L6 into the wall.
3. Only if a remote control is used: Pull cable L9 into the wall.
4. Pick up the adapter plate.
5. Insert the power cable and any existing cables L6 or L9 through the
adapter plate.
6. Screw on the adapter plate.
7. Align the adapter plate with a spirit level and screw it into place.
8. Attach the supplied caps to the screw heads.
empty
Surface installation
1. Pick up the adapter plate.
2. Screw on the adapter plate.
3. Align the adapter plate with a spirit level and screw it into place.
4. Attach the supplied caps to the screw heads.
empty
Further installation steps
➢ Perform the remaining installation steps as described in the Chapter
Adaptation of the remaining installation steps due to replacement of an
old unit:
● The drilling template is not required.
● The wall adapter is fastened using the three M8x16 screws and three
8.4mm washers supplied.
CAUTION
Do not install the cables for Remote Timer L6 and power cables in the
same conduit.
Содержание Heliodent plus
Страница 1: ...NNKOMNM ebiflabkqmirp k W b f f HELIODENT PLUS 7mA DC 7mA DC 70kV 60kV S 008...
Страница 2: ......