14 SIVISION 3
Sirona Dental Systems GmbH
14.3 SIROCAM 3 intraoral camera (extra equipment)
Operating Instructions C4
+
59 06 313 D 3382
74
D 3382.201.01.13.02
14.3
SIROCAM 3 intraoral camera (extra equipment)
The SIROCAM 3 intraoral camera combined with SIVI-
SION 3 is a perfect system for effective communication
with the patient.
The camera images cannot be used for diagnosis!
The SIROCAM 3 intraoral camera is a sensitive optical
instrument and must therefore always be handled with
care.
The camera illuminates the tooth areas to be imaged
with 7 white-light LEDs (1).
These are arranged around the entrance prism (2).
The LEDs and the lens are protected by a scratch-proof
antiglare glass cover.
CAUTION
* = Intermittent camera operation:
t1
= 1 min on,
t2
= 3 min off.
The camera heats up during operation.
To prevent overheating of the camera, the camera must
be put down for a time interval
t2
= 3 min after an operat-
ing time of
t1
= 1 min.
The camera will heat up even when it is placed in its
holder, as one of the states “black screen”, “white
screen” or “quad image” is always active.
The slide switch (3) is used for presetting the depth of
focus.
The camera button (4) is intended for controlling differ-
ent SIVISION 3 functions.
NOTICE
i
For camera operation and further information, please re-
fer to the operating instructions of the SIROCAM 3.
NOTICE
i
The camera handpiece contains data making it possible
to restore the factory-set color definitions regardless of
the dental treatment center where the camera is used.
Thus the camera can be used at different dental treat-
ment centers. To customize the color settings, you can
perform a white balance as long as the camera is active,
(see Mode dialog, section 11.11).
1
2
3
t
1
/
t
2
*
4