SPRAYVIT
If the SPRAYVIT is the only instrument that has been
picked up, you may switch the instrument light on or off.
You can also set the brightness.
The instrument light switches on when you activate the
SPRAYVIT (if preset).
The instrument light switches off after a time lag of 10s
when the SPRAYVIT is no longer activated.
When you place the SPRAYVIT back in its holder, the
instrument light switches off immediately.
If the SPRAYVIT is activated together with another
instrument, the SPRAYVIT light is not switched on.
DANGER
After changing hoses, press the water button of the
SPRAYVIT repeatedly until water comes out! Only
then can you begin treatment.
9.7
Saving the instrument programs
The factory-set instrument programs can be changed
individually by user
A
and user
B
.
Select the corresponding user, i.e.
A
or
B
, pick up an
instrument and set it according to your wishes.
Example for electric motor:
•
40,000 rpm
•
CW rotation
•
with cooling spray
•
with instrument light
To save the settings, press key
A/B
for some time
(approx. 3 s). An audible signal sounds. These settings
are then activated whenever the instrument is picked up.
You cannot switch between the two users after pick-
ing up an instrument.
NOTICE
i
If counterclockwise rotation was programmed, the motor
is reset to clockwise rotation after you switch the treat-
ment center off and back on.