Sirona Dental Systems GmbH
9 Dentist element and instrument functions
Operating Instructions C4
+
9.10 Satelec Mini L.E.D. curing light
59 06 313 D 3382
D 3382.201.01.13.02
47
Additional safety information
Any condensation occurring in the handpiece of the Mini
L.E.D. may cause impairments (e.g. fogging of the
L.E.D.). If the handpiece is moved from a cool environ-
ment to a warm room, always wait for it to reach room
temperature before putting it into operation.
Curing lights must not be used on persons who are suf-
fering from or have in the past been afflicted by photobi-
ological reactions (including solar urticaria and erythro-
poetic porphyria). Nor should they be used on persons
currently being treated with any medicine which
increases one’s sensitivity to light (including methox-
salene and chlorotetracycline).
Any persons who have in the past suffered from retinal
or eye lens disorders or have undergone eye surgery,
especially for gray cataracts, must consult their medical
eye specialist before using the light or undergoing treat-
ment with the Mini L.E.D. Caution is advisable even if the
patient gives his or her consent, since the light intensity
can cause accidents. It is especially advisable to always
wear appropriate protective goggles (UV filter).
Installation
1.
Connect the supply cable to the treatment center.
2.
Screw the handpiece on.
3.
Attach the sterilized light guide to the handpiece.
Make sure to insert the light guide correctly.
The light guide engages with a “click”.
4.
Slip the glare shield onto the light guide. It protects
your eyes against reflected light.
1
2
3
4