Sirona Dental Systems GmbH
10 Operation light, tumbler filling, cuspidor flushing, purge
Operating Instructions C4
+
10.4 Cuspidor flushing
59 06 313 D 3382
D 3382.201.01.13.02
53
CAUTION
• The tumbler must not be transparent.
• The tumbler must always be placed in the center of the
area provided for it.
• If an empty tumbler is standing below the tumbler filler
when the treatment center is switched on with the main
switch, the tumbler will not be filled automatically. To ac-
tivate automatic tumbler filling, remove the tumbler brief-
ly and then put it back again.
• The Mode dialog “CUP FILL ON S YES – NO” must be
set in such a way that “NO” flashes.
• Hydrocolloid connection is not possible if tumbler filling
with automatic sensor control is installed.
10.4
Cuspidor flushing
When you press this key briefly, the flushing function is
switched ON and the cuspidor is flushed for a preset
period of time.
When you press the key again while the cuspidor is
being flushed, flushing stops.
Only for cuspidor flushing key of the dentist ele-
ment:
Pressing this key for > 3 seconds displays the settings
dialog
You can set a flushing time between 20 and 420 sec-
onds (in steps) with the –
/
+ keys.
A
CUP FILL ON S
YES – NO
BOWL FLUSH TIMER
A