10 Operation light, tumbler filling, cuspidor flushing, purge
Sirona Dental Systems GmbH
10.5 Purge / Auto Purge function (purging the water paths)
Operating Instructions C4
+
59 06 313 D 3382
56
D 3382.201.01.13.02
If you press quick setting key 1 briefly (< 3 s), the follow-
ing prompt will appear if all instruments are in their hold-
ers:
Remove the instruments to be purged from the dentist
and assistant elements and insert them in the adapters
of the sanitation bowl as far as they will go.
CAUTION
Be careful not to snap off the instrument hoses when in-
serting the instruments
.
The maximum water flow must be set on the instru-
ments.
Press quick setting key 1 again (< 3 s), the instruments
are then purged one after the other.
The purging time in seconds set previously in the Mode
dialog (11.14) runs down for each instrument individu-
ally.
An audible signal sounds after the last instrument has
been flushed.
You now can continue in either of the following ways:
NOTICE
i
When removing the sanitation bowl, please notice that it
is filled with water.
A
SELECT ITEMS
FOR PURGING
SELECT ITEMS
FOR PURGING
1
wrong
right
•
Leave the instruments in the sanitation bowl
:
The treatment center can then be switched off.
The instruments will automatically be purged as
soon as the treatment center is switched on
again.
The APURGE function remains activated.
•
The instruments are returned to their holders:
The treatment center can be prepared for daily
practice operation
.
The APURGE function is completed.