Instruction Manual StackGuard 2 System
38
10198E/3
5.6
Replacing the purge air filter
Find the location of the purge air filter on the fold-out illustration in the
Appendix.
Action
1.
Switch the installation off (
2.
Detach the purge air hose (A) from the purge
air filter (B).
3.
Remove the two screws (C) and take off the
filter mount.
4.
Replace the purge air filter (B) with a new
one and fasten it in place with the filter
mount.
5.
Attach the purge air hose (A).
6.
Restart the installation.
7.
Enter the work done in the service log.