Instruction Manual StackGuard 2 System
10198E/3
21
The display...
Means that…
So you should …
**** Fault ****
...
…the functions of the ring
pipe control system
continue
unchanged
(blower is on, valves are
open, heaters are
activated, monitors are
activated).
…the current output is set
to 0 mA.
…if configured accordingly,
a relay contact is switched
…the automatic checking
function is deactivated.
…if the cause of the fault
no longer occurs, the fault
report will be extinguished.
-
...try to narrow down
the malfunction
(
Section 6.1).
** Prior.fault**
...
…the ring pipe is
switched
off
.
…the current ouptut is set
to 0 mA.
…if so configures, a relay
contact is switched.
…the automatic checking
function is deactivated.
…if the cause of the fault
no longer occurs, the fault
report will
not
be
extinguished.
-
...try to narrow down
the malfunction
(
Section 6.1).