Instruction Manual StackGuard 2 System
16
10198E/3
Item Name
Function
1
Main switch
Switches power to the entire installation on and
off.
2
Fasteners
The front cover (3) is opened by turning the two
fasteners.
3
Front cover
Protects the control unit against external effects.
4
Operating panel The system is operated with this panel (
Section
4.2).
Checking and
safety elements
5
Earth-leakage
circuit breaker
Protects the heater and the main blower.
6
Excess-current
circuit breaker
Protects the heater. The operating levers are
connected with a link, so that all levers are
operated simultaneously when the breaker trips.
7
Heat pack
Protects the main blower. The heat pack is made
up of these elements:
Rotary switch (for setting the main blower’s
rated current)
Cutout (for switching off the heat pack for
test purposes
Reset button
Do not alter rotary switch (A)!
8
Test sockets
4 .. 20mA
The current output signal can be measured at these
test sockets.
A short-circuit jumper connects the two test
sockets (4 .. 20mA
)
with each other, thus creating
a closed current loop.
9
RS-232
interface
RS-232 interface for diagnostic purposes
10
Expert button
Sealable button. Pressing this button activates the
expert mode.
This function is available to authorized persons
only!
11
Guard door
The guard door protects personnel against lethal
voltages.
Before opening the door, always make absolutely
sure that no lethal electric voltages are still
connected to the instrument!
Содержание StackGuard 2
Страница 2: ...SIGRIST PHOTOMETER AG technical details subject to change 08 2005...
Страница 6: ......
Страница 58: ...Instruction Manual StackGuard 2 System 52 10198E 3...
Страница 60: ...Instruction Manual StackGuard 2 System 54 10198E 3...
Страница 62: ...Instruction Manual StackGuard 2 System 56 10198E 3 W warning 8 warning signal 19 Z zero point reading 32...