Instruction Manual StackGuard 2 System
10198E/3
25
Action
Display
Remarks
6.
/
Change
number
/
Change
menu point
Limit
> 1 <
Limit monitor
(active "1", inactive "0")
Prio Fault Warn.
> 0 0 0 <
Active in the case of:
- prio fault
- fault
- warning
(active "1", inactive "0")
Serv. Check
> 0 0 <
Active when:
- instrument in service mode
- sensor check running
(active "1", inactive "0")
Dig.input
> 1 _ <
Active when:
- signal at control input 1
- signal at control input 2
(active -> number; inactive
"_")
Inversed
> 0 <
Invert relay function
(active "1", inactive “0")
7.
Relay N
>LI al se DI in<
Select additional function
(point 5) or terminate with
point 8
8.
> Relay N <
LI al se DI in
Confirm selection
9.
> Upper limit <
1.000 PLA
The limit is monitored only
when the limit function is
activated for the relay in
question
10.
Upper limit
> 1.000 PLA <
Activate editing mode
11.
/
Change
number
/
Change
place
Upper limit
> 2.000 PLA <
Set upper threshold
12.
> Upper limit <
2.000 PLA
Confirm entry
13.
> Lower limit <
0.900 PLA
14.
Lower limit
> 0.900 PLA <
Activate editing mode
15.
/
Change
number
/
Change
place
Lower limit
> 1.800 PLA <
Set lower threshold
Содержание StackGuard 2
Страница 2: ...SIGRIST PHOTOMETER AG technical details subject to change 08 2005...
Страница 6: ......
Страница 58: ...Instruction Manual StackGuard 2 System 52 10198E 3...
Страница 60: ...Instruction Manual StackGuard 2 System 54 10198E 3...
Страница 62: ...Instruction Manual StackGuard 2 System 56 10198E 3 W warning 8 warning signal 19 Z zero point reading 32...