Instruction Manual StackGuard 2 System
30
10198E/3
5.2
Checking the zero and reference points
The zero and reference point check is for ongoing quality assurance in
accordance with EN14181 (QAL3) and should be performed when the unit is
warm from operation!
Make sure that the checking rod is in the photometer only for a short time,
since the heat in the measuring cell changes the checking rod value and may
result in an incorrect adjustment.
When working on parts of the system that are not insulated, it is imperative
that temperature-resistant gloves are worn!
Action
Display (example)
Remarks
1.
Activate service
mode
* STEP OPERATION *
* *
Section 4.2.4
2.
2 x
* RECALIBR. *
* *
3.
1 x
check the
adjustment
nominal value.
Nominal value
must correspond
to the value on
the checking rod!
- Adjust. Nominal
-
7.52 PLA
If the nominal value does not
match the value on the
checking rod, contact your
service partner.
4.
1 x
- Reading <
0.0010
5.
Check the glasses on the checking rod for
cleanliness. Clean them if dirty with a
non-fluffy cotton cloth.
6.
Remove the clamping strips from the
insulation shells.
When working on parts of the system
that are not insulated, it is imperative that
temperature-resistant gloves are worn!
Zero and reference
point check
Содержание StackGuard 2
Страница 2: ...SIGRIST PHOTOMETER AG technical details subject to change 08 2005...
Страница 6: ......
Страница 58: ...Instruction Manual StackGuard 2 System 52 10198E 3...
Страница 60: ...Instruction Manual StackGuard 2 System 54 10198E 3...
Страница 62: ...Instruction Manual StackGuard 2 System 56 10198E 3 W warning 8 warning signal 19 Z zero point reading 32...