Instruction Manual StackGuard 2 System
10198E/3
17
4.2
Keypad and display
Figure 11: Operating elements and display.
/
Change from one menu line to another
Change numbers in editing mode (see below)
/
Change from one function to another in a menu line
Change function values or shift a number’s decimal point in
editing mode (see below)
+
Press both keys simultaneously to return to normal operation
Activate editing mode (display shows > <)
Enter the setting
4.2.1
Standby operation
The StackGuard 2 System is in standby mode whenever the main switch
(
Section 4.1) is in the "OFF" position. The system then switches to the
following state:
Photometer and SIGAR2 are switched off
No reading is available.
Main blower, purge air blower and heater are switched off.
Valves are closed
The ring pipe cools down gradually.
As the system cools down the measurement gas can condense, which can
cause corrosion in the ring pipe. Therefore the installation should be switched
off as infrequently as possible.
Key functions
Enter key
Up key
Right key
Reading
Down key
Left key
Unit
Measuring range