Instruction Manual StackGuard 2 System
42
10198E/3
Message
W F
P Means
Possible causes
T_max.entr.
x The temperature of
the supply pipe has
risen above the
maximum
admissible value.
-
medium too hot in
sample duct
-
malfunction of the
heater control
system
-
ring pipe clogged
T_min.return
x The temperature of
the return pipe has
dropped below the
minimum
admissible value.
-
heater failure
-
problem with ring
pipe (leakage)
T_max.return
x The temperature of
the return pipe has
risen above the
maximum
admissible value.
-
malfunction of the
heater control
system
-
ring pipe clogged
T_max.suppl.
x The third (optional)
temperature
monitor reports
overtemperature.
-
malfunction of the
heater control
system
-
ring pipe clogged
Rinse air
x Proper supply of
the flow cell with
purge air is no
longer assured.
-
purge air flow too
low
-
purge air blower
failure
-
purge air filter
clogged
Entr.val.op
x No check-back
signal from the
supply valve
indicating that it is
open.
-
defective valve
Entr.val.cl
x No check-back
signal from the
supply valve
indicating that it is
closed.
-
defective valve
Ret.val.op
x No check-back
signal from the
return valve
indicating that it is
open.
-
defective valve
Ret.val.cl
x No check-back
signal from the
return valve
indicating that it is
closed.
-
defective valve
Содержание StackGuard 2
Страница 2: ...SIGRIST PHOTOMETER AG technical details subject to change 08 2005...
Страница 6: ......
Страница 58: ...Instruction Manual StackGuard 2 System 52 10198E 3...
Страница 60: ...Instruction Manual StackGuard 2 System 54 10198E 3...
Страница 62: ...Instruction Manual StackGuard 2 System 56 10198E 3 W warning 8 warning signal 19 Z zero point reading 32...