Servicing
Instruction Manual AquaScat S
13798E/3
81
WORKSTEP
ADDITIONAL INFO / IMAGES
4.
Place the cover back onto the sensor head and
fasten in place with the two screws.
Tighten the screws carefully as
plastic threads are used.
5.
Clean the sapphire windows and the surface
where they are located with alcohol.
6.
Carry out recalibration according to Section
9.6.
7.
Install the instrument according to Section 4
and put it back into operation.