Instruction Manual AquaScat S
Mounting
32
13798E/3
4.7
Mounting the submerge tube version with special cable
The submerge tube version with special cable can only be used together with a SICON.
The length of the special cable must be selected so that the AquaScat S is always completely
immersed in the water (well).
The following points must be observed when mounting the submerge tube:
The AquaScat S must be protected against exposure to light.
The water to be measured must be free from air bubbles.
There must be no turbulence in the water in the immediate vicinity of the immersion
probe (e.g. caused by suction pumps).
The immersion zone of the probe should be positioned so that there is no risk of dam-
ages to the probe as a result of collisions.
The immersion depth of the probe must be selected so there is sufficient clearance to
the bottom (sludge and sediment on the bottom).
Check that the cable gland on the probe is securely mounted (protection against water
penetration).