Decommissioning/Storage
Instruction Manual AquaScat S
13798E/3
101
12
Decommissioning/Storage
12.1
Decommissioning the photometer
The aim of decommissioning is to prepare the individual components of the system properly
for storage.
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage and remove the
connector from the AquaScat S.
2.
Remove the AquaScat S from the line accord-
ing to Section 4.
3.
Clean and dry the AquaScat S.
4.
Remove the electrical connections from the
Conn-R junction box and the SICON (M) (if
present).
Section 5
5.
Remove and pack all components.
12.2
Storing the photometer
There are no special requirements for storing the instruments. However, please note the
following information:
The system contains electronic components. Storage for such components must fulfill
the usual conditions. It is important to note that the storage temperature must be be-
tween -20 and +50 °C.
All components that come into contact with the sample during operation have to be dry
and clean for a long time before being put into storage.
The measuring equipment with all of the accessory parts must be protected against
weather factors, condensing humidity and aggressive gases during storage.