Instruction Manual AquaScat S
Mounting
24
13798E/3
4
Mounting
4.1
Basic information for mounting the AquaScat S
Figure 5: Mounting directions
The following points must be observed when mounting the AquaScat S:
The photometer should be mounted in positions where there is a uniform flow. This is
usually the case in the standpipe (pos. 1). The photometer can also be installed in pipes
that run horizontally (pos. 3).
Do not mount the photometer in the downpipe (pos. 2).
The photometer must be mounted in the line at least 0.5 meters away from inspection
glasses and other unwanted light sources.
The angled edge (pos. 4) of the sensor head must always be positioned against the flow
direction (arrow) (pos. 1).
Do not mount the photometer in locations where bubbles may form.
Do not mount the photometer downstream of pressure reduction, as cavitation may
form here.