Commissioning
Instruction Manual AquaScat S
13798E/3
51
WORKSTEP
ADDITIONAL INFO / IMAGES
4.2: Change the parameters as required.
To do this, enter a number after “
=
” and con-
firm with the Enter button (e.g. 0 for German
or 1 for English).
Language:
0: German, 1: English
Setting the operating language
Current from:
0.000
Setting the lower value of the cur-
rent range
Current to:
1000
Setting the upper value of the cur-
rent range
Limits, Mode:
0: Inactive, 1: Exceeded, 2: Under-
shot. Default = 0
Limits, Upper limit:
1.000. Enter
the limit with the upper threshold
value.
Limits, Lower limit:
0.900. Enter
the limit with the lower threshold
value.
Integration:
Integration time of the
measuring value from 0 .. 255 sec-
onds. Default = 5 s
Output 1, Invert:
0: No, 1: Yes
Invert: Inverts output 1 (alarm)
Output 2, Invert:
0: No, 1: Yes
Invert: Inverts output 2 (limit)
Factory set.:
0: No, 1: Yes
Yes
loads the factory setting.
4.3: Close the file after making your entries.
Changes are saved automatically.
The window of the removable disk closes and
then reopens after a short time.
4.4: If configuration was successful, an addi-
tional file
Config.OK
appears.
If configuration was unsuccessful, an additional
file
Config.ERR
appears.
The configuration can fail if values outside the
permitted range are entered or if text fields
have been deleted or changed.
In the event of an incorrect configuration, the
Config.txt
file is regenerated automatically
and the procedure can be repeated.
5.
Carry out zero calibration on the AquaScat ac-
cording to Section 9.6.4.
6.
Remove the USB connector, replace the desic-
cant according to Section 9.2and reattach the
cover.
If faults occur, please refer to
the Reference Manual.