Instruction Manual AquaScat S
Commissioning
48
13798E/3
6
Commissioning
The initial start-up of the web user interface via the optional WLAN interface is described in
the Reference Manual.
6.1
Commissioning with Conn-R junction box and SICON-C
Proceed with the initial start-up in accordance with the following table:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Ensure that all components are correctly
mounted and connected.
2.
2.1: Establish the service voltage to the Conn-R
junction box (and thus to the photometer).
2.2: Plug the SICON-C into the Conn-R junc-
tion box according to Section 5.2.2 and wait
until the welcome screen appears in the dis-
play.
The factory setting language is English.
2.3: The instrument is ready for measurement.
3.
Set the language.
Section 8.1
4.
Set the limits.
Section 8.3
5.
Set the current outputs.
Section 8.2
6.
Set the outputs.
Section 8.4
7.
Enter the access code.
Section 8.9
8.
Carry out recalibration.
Section 9.6
9.
Back up the configured data.
Section 8.10