Instruction Manual AquaScat S
Mounting
28
13798E/3
4.4.2
Extractable assembly: Mounting the probe in a horizontal pipe
On horizontal pipes, the handle on the probe must be turned by 90°. This can be done as
follows:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Loosen the screw (X) so that the handle can be
turned easily.
2.
Align the handle to the marking on the probe
according to the current installation situation.
The following three markings are found on the
handle:
A: Vertical line, flow from below
B: Horizontal line, flow from right
C: Horizontal line, flow from left
3.
Fasten the handle in place by tightening the
screw (X) on the probe.
4.
Rotate the complete assembly until the hose
connection (arrow) is pointing downwards.
5.
Rotate the handle into the horizontal position
and ensure the screw (X) is finally positioned at
the top.
6.
Reinstall the probe into the extractable assem-
bly as detailed in Section 9.5(step 10 to 17).
,