Instruction Manual AquaScat S
Commissioning
50
13798E/3
6.3
Commissioning without SICON (M)
Proceed with the initial start-up in accordance with the following table:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Ensure that the photometer is correctly mount-
ed and connected.
2.
Establish the service voltage.
3.
Connect the AquaScat S to the PC as follows:
3.1: Remove the cover on the AquaScat S.
3.2: Connect the USB cable to the AquaScat
(arrow) and connect to the PC. The AquaScat S
is automatically detected as a removable disk
(Windows operating system).
3.3: Depending on the Windows configuration,
either a window for the removable disk is
opened automatically or Windows Explorer
must be started and the removable disk
opened manually.
The following files are shown:
Info.txt
: Information on the AquaScat S, such
as serial number, software version and operat-
ing hours.
Config.txt
: Configuration file.
Zero.txt
: File used for initiating zero calibra-
tion.
4.
Carry out configuration as follows:
4.1: Open the
Config.txt
file.