Instruction Manual AquaScat S
Servicing
78
13798E/3
9.3
Replacing the desiccant on the submerge tube version
with special cable
The following procedure describes how the desiccant is changed in the AquaScat S when the
submerge version with special cable is used:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Interrupt the service voltage to the AquaScat S.
2.
Loosen the cable gland (arrow) and slide the
submerge tube housing off the AquaScat S
(figure below).
3.
Remove the M12 connector from the
AquaScat S.
4.
Unscrew the cover on the AquaScat S.
The cable connection between the
cover and instrument remains in place.
5.
Replace the old desiccant bag with a new one.