Servicing
Instruction Manual AquaScat S
13798E/3
79
WORKSTEP
ADDITIONAL INFO / IMAGES
6.
Replace the 35 x 1.5 gasket on the cover.
In order for the cable between the cover
and instrument to remain connected, the gas-
ket should be removed and attached from the
outside.
7.
Screw the cover back onto the AquaScat S.
In doing so, ensure that the cables are not
twisted before screwing the cover into place.
8.
Coat the O-ring in the submerge tube housing
with food-grade grease (arrow).
9.
Refasten the M12 connector on the
AquaScat S.
10. Push the submerge tube over the AquaScat S
up to the stop.
11. Pull the cable taught, then tighten the cable
gland (arrow).