SIGLENT
36 SSG5000X Programming Guide
Range
Frequency between 9 kHz ~ 100 kHz: -110 dBm ~ 9 dBm
Frequency between 100 kHz ~ 1 MHz: -110 dBm ~ 15 dBm
Frequency above 1 MHz: -110 dBm ~ 20 dBm
Return
Float, unit: dBm
Default
-130 dBm
Menu
LEVEL > Level
Example
POWer 2
3.6.2.3
Level Offset ([:SOURce]:POWer:OFFSet)
Command
Format
[:SOURce]:POWer:OFFSet <power>
[:SOURce]:POWer:OFFSet?
Instruction
Set the RF offset level of the RF output
Get the RF offset level of the RF output
Parameter
Type
Float
Parameter
Range
-100 dB ~ 100 dB
Return
Float, unit: dB
Default
0 dB
Menu
LEVEL > Level Offset
Example
POWer:OFFSet 2
3.6.2.4
ALC State ([:SOURce]:POWer:ALC)
Command
Format
[:SOURce]:POWer:ALC ON|OFF|AUTO
[:SOURce]:POWer:ALC?
Instruction
Activate/deactivate automatic level control.
Query ALC state
Parameter
Type
Enumeration
Parameter
Range
ON|OFF|AUTO
ON
Internal level control is permanently activated.
OFF
Internal level control is deactivated; Sample & Hold mode is activated.
Содержание SSG5000X Series
Страница 1: ...SSG5000X Series Signal Generator Programming Guide PG0805X E01A...
Страница 14: ...SIGLENT 14 SSG5000X Programming Guide...
Страница 131: ...SIGLENT SSG5000X Programming Guide 131...
Страница 152: ...SIGLENT 152 SSG5000X Programming Guide To close the normal telnet window type Quit and press Enter...