SIGLENT
SSG5000X Programming Guide 67
Example
PULM:PER 220 us
3.6.8.7
Pulse Width ([:SOURce]:PULM:WIDTh)
Command
Format
[:SOURce]:PULM:WIDTh
<value>
[:SOURce]:PULM:WIDTh?
Instruction
Set the width of the generated pulse
Get the width of the generated pulse
Parameter
Type
Float, unit: ns, us, ms, s
Parameter
Range
20 ns ~ 300 s
Return
Float, unit: s
Default
2 ms
Menu
PULSE > Pulse Width
Example
PULM:WIDT 33 us
3.6.8.8
Double Pulse Delay ([:SOURce]:PULM:DOUBle:DELay)
Command
Format
[:SOURce]:PULM:DOUBle:DELay
<value>
[:SOURce]:PULM:DOUBle:DELay
?
Instruction
Set the delay from the start of the first pulse to the start of the second pulse
Get the delay from the start of the first pulse to the start of the second pulse
Parameter
Type
Float, unit: ns, us, ms, s
Parameter
Range
20 ns ~ 300 s
Return
Float, unit: s
Default
4 ms
Menu
PULSE > Double Pulse Delay
Example
:PULM:DOUBle:DELay 2 ms
3.6.8.9
#2 Width ([:SOURce]:PULM:DOUBle:WIDTh)
Command
Format
[:SOURce]:PULM:DOUBle:WIDTh <time>
[:SOURce]:PULM:DOUBle:WIDTh
?
Содержание SSG5000X Series
Страница 1: ...SSG5000X Series Signal Generator Programming Guide PG0805X E01A...
Страница 14: ...SIGLENT 14 SSG5000X Programming Guide...
Страница 131: ...SIGLENT SSG5000X Programming Guide 131...
Страница 152: ...SIGLENT 152 SSG5000X Programming Guide To close the normal telnet window type Quit and press Enter...