72 SSG5000X Programming Guide
Instruction
Select the trigger mode for pulse modulation
Get the trigger mode for pulse modulation
Parameter
Type
Enumeration
Parameter
Range
AUTO| KEY|EXTernal|EGATe
Return
Enumeration
Default
AUTO
Menu
PULSE > Pulse Trigger
Example
PULM:TRIG:MODE EXTernal
3.6.8.20
Trig Polarity
([:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity)
Command
Format
[:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity
NORMal|INVerted
[:SOURce]:PULM:TRIGger:EXTernal:GATE:POLarity?
Instruction
Select the polarity of the gate signal
Get the polarity of the gate signal
Parameter
Type
Enumeration
Parameter
Range
NORMal|INVerted
Return
Enumeration
Default
NORMal
Menu
PULSE > Pulse Polarity
Example
PULM:TRIG:EXT:GATE:POL NORMal
3.6.8.21
Trig Delay ([:SOURce]:PULM:DELay)
Command
Format
[:SOURce]:PULM:DELay
<value>
[:SOURce]:PULM:DELay
?
Instruction
Set the pulse delay
Get the pulse delay
Parameter
Type
Float, unit: ns, us, ms, s
Parameter 140 ns ~ 300 s