SIGLENT
110 SSG5000X Programming Guide
Default
-123.27 dBm
Menu
ARB > ARB Setup > Real Time AWGN > Channel Power
Example
:RADio:ARB:NOISe:POWer:CHNOise 0 dBm
3.8.4.13
Total Noise Power ([:SOURce]:RADio:ARB:NOISe:POWer:
TONOise)
Command
Format
[:SOURce]:RADio:ARB:NOISe:POWer:TONOise <power>
[:SOURce]:RADio:ARB:NOISe:POWer:TONOise?
Instruction
This command sets the total noise power of the Real Time AWGN.
Parameter
Type
Float
Parameter
Range
According to the total power limit
Return
Float
Default
-122.76 dBm
Menu
ARB > ARB Setup > Real Time AWGN > Noise Power
Example
:RADio:ARB:NOISe:POWer:TONOise 0 dBm
3.8.4.14
Carrier To Noise Ratio Format
([:SOURce]:RADio:ARB:NOISe:CN:FORMat)
Command
Format
[:SOURce]:RADio:ARB:NOISe:CN:FORMat CARRier|BIT
[:SOURce]:RADio:ARB:NOISe:CN:FORMat?
Instruction
This command sets the carrier to noise ratio format of the Real Time AWGN.
Parameter
Type
Enumeration
Parameter
Range
CARRier|BIT
Return
Enumeration
Default
CARRier
Menu
ARB > ARB Setup > Real Time AWGN > Carrier To Noise Ratio Format
Example
:RADio:ARB:NOISe:CN:FORMat BIT