SIGLENT
SSG5000X Programming Guide 117
Return
Enumeration
Default
None
Menu
ARB > Marker Utilities > Pulse/RF Blank
Example
:RADio:ARB:MDEStination:PULSe Marker2
3.8.4.29
Clipping ([:SOURce]:RADio:ARB:CLIPping)
Command
Format
[:SOURce]:RADio:ARB:CLIPping "<file name>",IJQ|IORQ,<val>[,<val>]
Instruction
This command sets the clipping level of the selected waveform segment to
a percentage of its highest peak
Parameter
Type
String, Enumeration, Float, Float
Parameter
Range
None, IJQ|IORQ, 0.01 ~ 1, 0.01 ~ 1
Return
None
Default
None
Menu
ARB > Waveform Utilities > Clipping
Example
:RADio:ARB:CLIPping “SINE_WAVE”,IJQ,0.75
3.8.4.30
Scaling ([:SOURce]:RADio:ARB:SCaling)
Command
Format
[:SOURce]:RADio:ARB:SCaling "<file name>",<val>
Instruction
This command scales the designated "<file_name>" waveform file
Parameter
Type
String, Float
Parameter
Range
None, 0.01 ~ 1
Return
None
Default
None
Menu
ARB > Waveform Utilities > Apply to Waveform
Example
:RADio:ARB:SCaling “SINE_WAVE”,0.75
Содержание SSG5000X Series
Страница 1: ...SSG5000X Series Signal Generator Programming Guide PG0805X E01A...
Страница 14: ...SIGLENT 14 SSG5000X Programming Guide...
Страница 131: ...SIGLENT SSG5000X Programming Guide 131...
Страница 152: ...SIGLENT 152 SSG5000X Programming Guide To close the normal telnet window type Quit and press Enter...