100 SSG5000X Programming Guide
Return
Boolean
Default
0
Menu
Multitone > Single Side
Example
:RADio:MTONe:ARB:SETup:TABLe:SINGle 1
3.8.2.4
Sample Rate
([:SOURce]:RADio:MTONe:ARB:SCLock:RATE)
Command
Format
[:SOURce]:RADio:MTONe:ARB:SCLock:RATE <val>
[:SOURce]:RADio:MTONe:ARB:SCLock:RATE?
Instruction
This command sets the sample clock rate for the Multitone modulation
format
Parameter
Type
Float, unit: Hz, kHz, MHz, GHz, Default “Hz”
Parameter
Range
500 Hz ~ 240 MHz
Return
Float
Default
2 MHz
Menu
Multitone > Sample Rate
Example
:RADio:MTONe:ARB:SCLock:RATE 5000000
3.8.2.5
Freq Spacing
([:SOURce]:RADio:MTONe:ARB:SETup:TABLe:FSPacing)
Command
Format
[:SOURce]:RADio:MTONe:ARB:SETup:TABLe:FSPacing <val>
[:SOURce]:RADio:MTONe:ARB:SETup:TABLe:FSPacing?
Instruction
This command sets the frequency spacing between the total tones
Parameter
Type
Float
Parameter
Range
500 Hz ~ 120 MHz
Return
Float
Default
1 MHz
Menu
Multitone > Sample Rate
Example
RADio:MTONe:ARB:SETup:TABLe:FSPacing 2000000