90 SSG5000X Programming Guide
Type
Parameter
Range
PN7|PN9|PN15|PN23|USER
Return
Enumeration
Default
PN7
Menu
Custom > Data Source > Data Setup
Example
:RADio:CUSTom:DATA PN9
3.8.1.3
Symbol Rate ([:SOURce]:RADio:CUSTom:SRATe)
Command
Format
[:SOURce]:RADio:CUSTom:SRATe <val>
[:SOURce]:RADio:CUSTom:SRATe?
Instruction
This command sets the transmission symbol rate. Symbol rate is the bit rate
divided by the bits per symbol. A change in the symbol rate affects the bit
rate
Parameter
Type
Float
Parameter
Range
500 Sps ~ 120 MSps
Return
Float
Default
1 MSps
Menu
Custom > Data Source > Symbol Rate
Example
:RADio:CUSTom:SRATe 2000000
3.8.1.4
Symbol Length ([:SOURce]:RADio:CUSTom:SLENgth)
Command
Format
[:SOURce]:RADio:CUSTom:SLENgth <val>
[:SOURce]:RADio:CUSTom:SLENgth?
Instruction
This command sets the transmission symbol length
Parameter
Type
Integer
Parameter
Range
100 ~ 100000
Return
Integer
Default
512
Menu
Custom > Data Source > Symbol Length