M
ESSAGES AND
D
IAGNOSIS
Error
No.
Data and operating
errors
- Meaning
Explanation
79
Number of weight
ranges not allowed
The number of the weighing range is incorrect. Enter 1, 2 or 3.
81 Relationship
between
weighing ranges
implausible
The definition for weighing ranges is incorrect in DR 3 (Max-
and Min-values).
On a multi-resolution scale, the Min-value of the next higher
range must be equal to the Max-value of the range
underneath. The Max-value must be greater than the Min-
value.
On a multi-range scale, the Min- as well as the Max-values
must be increasing sequential ranges and the Max-value of a
range must always be greater than the Min-value.
Check parameter definitions.
82 Numeral
step
not
plausible
At least one numeral step is incorrect in DR 3.
The following numeral steps are permitted as numeral steps
for the three weighing ranges and the distribution value: 50,
20, 10, 5, 2, 1, .., 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0,005, 0,002,
0,001, 0.0005, 0.0002, 0.0001.
The numeral step is not permitted to be to be less than one
6000th of the weighing range maximum value in verifiable
operation (technical regulation code OIML in DR 3).
The total numeral step may not be less than the numeral step
of weighing range 1 in verifiable operation (regulation code
OIML set in DR 3).
The numeral step of a weighing range may not be less than
the numeral step of the weighing range below it.
Check parameter definitions.
83 Filter
parameter
implausible
The filter parameter definition in DR 3 or DR 10 is not
plausible. Check the codes for the filter type (0..2) and the limit
frequency (0..9).
84 Characteristic
value
not plausible
The characteristic value in DR 3 is incorrect. Set the
parameter to a valid value (1, 2 or 4).
85
Averaging filter length
not permitted
The average value filter is defined incorrectly in DR 3. Set the
parameter for the depth of the average value filter to 250 at
the very most.
86 Adjustment
weight
error
The defined adjustment weights are incorrect. The weight
values must be in ascending order or 0 if not used.
87 Conversion
factor
error
The conversion factor from weight unit to large weight unit is
implausible.
88 Adjustment
zero
setting time too large
The adjustment time and zero setting time is too great (0…900
s).
89
Reserved
94
Adjustment weight too
low
The adjustment weight is too low in DR 3. Increase the
adjustment weight interval. The measurement values between
two sequential adjustment weights must be separated by a
space of at least 2% of the measurement range.