Error
No.
Data and operating
errors
- Meaning
Explanation
56
Log text too long
The log text with filled in parameters is too long or the end
character (EOT) is missing. Change the log text.
Ensure that in place of the corresponding print field code, the
respective number of characters for the field is set. The overall
length of the text filled with values may not exceed 160
characters (incl. printer control characters but without the
termination EOT).
57
Log text contains
fields that are not
permitted
Log text contains fields that are not permitted Change the log
text.
58
Log text contains
angle-brackets
Log text contains angle-brackets Remove the angle-brackets
from the log text.
62
Logging not possible
since the weight lies
outside of the
weighing range
Logs can only be output within the weighing range in verifiable
operation (code OIML in DR 3).
63
Logging output not
possible since the
SIWATOOL FTC
driver is active
The log can not be output because the SIWATOOL FTC_B is
active on the interface. Disconnect the PC and connect the
printer.
64
Logging output not
possible since the
printer is not ready
The log cannot be printed since the printer is not ready. Check
the printer.
65
Logging output not
possible since a print
job is still running
The log cannot be printed since another print job is still
running. Wait until the current print job is complete.
66
Logging not possible
since the MMC is full
Logging is not possible since the MMC is full. Switch MMCs.
67 Logging
task
cannot
be executed, since
MMC is defective or
non-existent
Logging is not possible since the MMC is defective. Switch
MMCs.
68
Log output not
possible since no data
is available.
Repeating the last log print-out is not possible since there has
been no previous print-out or the saved data is defective.
70
Logging output not
possible since the log
ID is not in the print
text
Log output is not possible since the content of the text does not
contain the log ID. Define the log ID in the text. This entry is
necessary if the verifiable weight is contained in the log text in
verifiable operation ((Technical regulation code OIML in DR 3).
76 Value
range
Adjustment digits
exceeded
The definition for the adjustment digits in DR 3 is too large.
Enter a smaller value. The highest permitted value is
16 777 215.
77 Regulation
code
not
permitted
The regulation code is incorrect. Enter the proper code.
The permitted entries are ---- or OIML.
78
Zero setting range >
4% or zero setting
range exceeded
The zero definition range or zero setting range is too large.
This is the case if you have defined a zero setting range of >
4% of the maximum weighing range, or > 20% of the zero
setting range in DR 3 in verifiable operation (technical
regulation code OIML in DR 3). The sum of the negative and
positive values apply in this instance. Reduce the zero
definition range or zero setting range.